2-wire transmitter for pH Type 202701/10 for redox Type 202701/20 The 2-wire transmitter is intended for linking a pH or redox combination electrode with plug con- nection to indicators/controllers with an active 4â20 mA input. On the output side the 2-wire transmitters have a connection for supply and standard signal. Zero and slope of pH combina- tion electrodes are adjusted at the indicator /controller. No calibration is required for redox elec- trodes. The 2-wire transmitter is screwed directly on to the electrode head of the combination elec- trodes. This circuit arrangement largely prevents interference from dirt, humidity, or electrical fields from live conductors. A conventional coaxial cable or a 2 wire cable with screen is sufficient as con- nection between the transmitter and the indicator. This permits trouble-free transmission over larger distances between the transmitter and the indicator. An isolated supply is recommended when operating the transmitters with a PLC. Type 202701 for pH The 2-wire transmitter converts the high-impedance signal of the pH electrode (up to 1000MOhm) into a standard 4â20 mA signal. Type 202702 for redox The 2-wire transmitter converts the signal of the redox electrode into a standard 4â20 mA signal.
|